Ubicación: Global + English
Global Global Algérie Français Algeria English Angola Português Angola English Argentina Español Argentina English Australia English Austria Deutsch Austria English Azerbaijan English Azerbaijan Русский Bahrain English Bangladesh English Belgium English Belgium Français Belgium Nederlands Brazil Português Brazil English Bulgaria български Bulgaria English Cameroon English Cameroon Français Canada English Canada Français Chile Español Chile English China 中文 China 日本語 China English Colombia Español Colombia English Croatia English Czech Republic Čeština Czech Republic English Denmark English Djibouti English Ecuador Español Ecuador English Egypt English Estonia English Ethiopia English Finland Suomi Finland English France Français France English Georgia English Georgia Русский Germany Deutsch Germany English Ghana English Greece Ελληνικά Greece English Guatemala Español Guatemala English Hong Kong, China English Hong Kong, China 中文 Hungary Magyar Hungary English India English Indonesia English Iraq English Ireland English Italy Italiano Italy English Ivory Coast Français Ivory Coast English Japan 日本語 Japan English Jordan English Kazakhstan Русский Kazakhstan English Kenya English Kuwait English Kyrgyzstan Русский Kyrgyzstan English Latvia English Lithuania English Malaysia English Mauritius English Mauritius français (Maurice) Mexico Español Mexico English Moldova Română Moldova English Mongolia English Morocco English Morocco Français Mozambique Português Mozambique English Netherlands Nederlands Netherlands English New Zealand English Nigeria English Norway English Oman English Pakistan English Paraguay Español Paraguay English Peru Español Peru English Philippines English Poland Polski Poland English Portugal Português Portugal English Qatar English Republic of Korea 한국어 Republic of Korea English Romania Română Romania English Saudi Arabia English Serbia Српски Serbia English Singapore English Slovakia English Slovenia English South Africa English Spain English Spain Español Sri Lanka English Sweden English Switzerland Deutsch Switzerland Français Switzerland Italiano Switzerland English Tanzania English Thailand ไทย Thailand English Togo English Togo Français Tunisia English Tunisia Français Türkiye Türkçe Türkiye English Turkmenistan Русский Turkmenistan English Ukraine Українська Ukraine English United Arab Emirates English United Kingdom English Uruguay Español Uruguay English USA English Uzbekistan English Uzbekistan Русский Vietnam Tiếng Việt Vietnam English

Determination of Total Atmospheric Ozone: Principles, Methods, and Applications

Determination of total atmospheric ozone (total column ozone) quantifies the integrated abundance of ozone in a vertical column of air, expressed in Dobson Units (DU), and provides the fundamental dataset for ozone-layer monitoring, UV-index forecasting, and climate research. Combining ground-based UV spectrophotometry, balloon-borne ozonesondes, and satellite remote sensing in accordance with WMO/GAW guidelines and ISO/EN ultraviolet photometric standards, accredited third-party laboratories and global monitoring networks deliver measurements with precision on the order of one percent.

Technical Overview

Total atmospheric ozone determination measures the integrated amount of ozone (O₃) contained in a vertical column extending from the Earth's surface to the top of the atmosphere. Approximately 90 percent of this ozone resides in the stratosphere between roughly 10 and 50 kilometers altitude, where it absorbs biologically harmful ultraviolet (UV) radiation, while the remaining tropospheric fraction acts as both a greenhouse gas and a key secondary pollutant in photochemical smog.

Column abundance is conventionally expressed in Dobson Units (DU): 1 DU corresponds to a layer of pure ozone 0.01 mm thick at standard temperature and pressure, equivalent to 2.687 × 10¹⁶ ozone molecules per square centimeter. Typical values range from about 230 DU in the tropics to more than 450 DU at high latitudes in spring, with a global average near 300 DU. Sustained column values below 220 DU over Antarctica define the phenomenon known as the "ozone hole."

The measurement principle is based on the strong absorption of solar UV radiation by ozone in the Hartley and Huggins bands (approximately 200–360 nm). Instruments compare the relative intensities of wavelength pairs—one strongly absorbed and one weakly absorbed—and, by applying the Beer–Lambert law with corrections for the solar zenith angle (optical air mass), Rayleigh scattering, aerosol attenuation, and the effective temperature of the ozone layer, retrieve the total column with precision of about 1 percent.

Systematic observations began with the Dobson spectrophotometer in the 1920s and expanded into a coordinated global network during the 1957–1958 International Geophysical Year. Today, measurements are harmonized under the World Meteorological Organization (WMO) Global Atmosphere Watch (GAW) programme, with validated data archived at the World Ozone and Ultraviolet Radiation Data Centre (WOUDC).

Testing Samples

Direct Solar Radiation at Ground Observatories

Clear-sky direct-sun observations are the primary measurement geometry. Solar UV radiation is collected through a quartz-window aiming head and analyzed at multiple wavelength pairs near local noon and throughout the day, providing the lowest-uncertainty determination of total column ozone.

Zenith-Sky and Moonlight Radiation

When clouds obscure the sun, or during polar night, instruments measure diffuse zenith-sky radiance or reflected moonlight. Empirical cloud correction tables and modified air-mass treatments are applied, with somewhat higher measurement uncertainty.

Balloon-Borne Air Column Profiles

Radiosonde flights carrying electrochemical ozonesondes sample ambient air continuously from the surface to burst altitudes near 30–35 km. Vertical integration of the recorded profile yields an independent column total and the vertical distribution of ozone.

Calibration Transfer Standards

Laboratory calibration samples include tungsten-halogen standard lamps, mercury discharge lamps for wavelength verification, primary reference spectrophotometers, and standard-reference photometers traceable to national metrology institutes, all used to maintain network-wide comparability.

Testing Items

Core measurands and associated quantities typically reported in a total ozone determination program include:

  • Total column ozone, reported in Dobson Units (DU)
  • Stratospheric and tropospheric partial ozone columns
  • Vertical ozone profile (from Umkehr, ozonesonde, or lidar observations)
  • Sulfur dioxide (SO₂) vertical column, co-retrieved as an interference correction
  • Aerosol optical depth and atmospheric turbidity
  • Effective ozone layer temperature used for absorption cross-section correction
  • Surface ozone concentration (µg/m³ or nmol/mol) determined by UV photometry per ISO 13964 and EN 14625
  • Instrument calibration parameters: extraterrestrial constants, standard-lamp response, wavelength accuracy, and stray-light ratio

Testing Methods

Dobson Spectrophotometry

The Dobson instrument uses a double-monochromator optical system to compare solar intensity at the A, B, C, and D wavelength pairs (305.5/325.4 nm through 317.6/339.9 nm). The AD double-pair technique cancels the influence of aerosols and thin clouds, and observations are reduced using WMO/GAW operational procedures for the Dobson network.

Brewer Spectrophotometry

The Brewer grating spectrophotometer measures direct-sun or zenith-sky radiances at five wavelengths between 306.3 and 320.1 nm. Routine direct-sun observations achieve repeatability near 1 percent, and the instrument simultaneously retrieves SO₂ column and spectral UV, following the WMO/GAW Brewer operations handbook.

Umkehr Profile Retrieval

By recording zenith-sky radiance ratios at solar zenith angles from about 60° to 90°, the Umkehr method retrieves coarse vertical ozone profiles in roughly 5–10 km layers, extending the utility of a single column instrument to profile information.

Balloon-Borne Electrochemical Ozonesondes

Electrochemical concentration cell (ECC) sondes pump ambient air through a potassium iodide solution, generating a current proportional to ozone. Profiles typically agree with column instruments to within 2–3 percent and are essential for validating satellite retrievals in the lower atmosphere.

Satellite Remote Sensing

Nadir-viewing sensors such as TOMS, SBUV/2, OMI, OMPS, and TROPOMI measure backscattered UV radiances and retrieve column ozone using differential optical absorption spectroscopy (DOAS) and backscatter-ultraviolet algorithms. Global maps are validated against ground-based Dobson, Brewer, and ozonesonde networks.

Surface Ozone Ultraviolet Photometry

For near-ground ozone concentration, continuous UV absorption photometers operating at 254 nm follow the reference methods of ISO 13964 and EN 14625, with traceable calibration transfer standards and periodic multipoint verification.

Calibration and Quality Assurance

Network coherence is maintained through a hierarchical system: primary Dobson standard triads, regional intercomparisons organized by WMO, standard-lamp and wavelength checks, portable reference spectroradiometers for site audits, and mandatory submission of quality-flagged data to WOUDC and complementary archives such as NDACC.

Applications

Total atmospheric ozone determination supports a broad range of scientific, regulatory, and operational functions:

  • Assessment of ozone-layer depletion and recovery under the Vienna Convention and Montreal Protocol
  • UV Index forecasting and public-health warnings regarding solar UV exposure
  • Climate research: ozone is an essential climate variable, and column data constrain radiative-forcing calculations and chemistry–climate model evaluation
  • Air quality management, including characterization of tropospheric ozone episodes and long-range pollutant transport
  • Validation and long-term stability monitoring of satellite ozone retrievals
  • Aviation and high-altitude operations, where elevated UV and ozone exposure requires assessment
  • Effects research on crops, materials degradation, and ecosystem UV dose

FAQ

What is a Dobson Unit (DU)?

One DU equals a pure ozone layer 0.01 mm thick at standard temperature and pressure, or 2.687 × 10¹⁶ molecules per square centimeter. A representative global column of 300 DU, compressed to sea-level pressure, would form a layer only about 3 mm thick.

How does total column ozone differ from surface ozone?

Total column ozone integrates the entire vertical atmosphere and is dominated by the stratospheric layer, reported in DU. Surface ozone is a near-ground concentration in µg/m³ or ppb, regulated under ambient air quality standards and measured by the UV photometric methods of ISO 13964 and EN 14625.

Can total ozone be measured under clouds or at night?

Yes. Zenith-sky observations with cloud corrections handle cloudy daytime conditions, moonlight observations extend coverage through polar night, and balloon-borne ozonesondes operate independently of daylight.

What accuracy can be expected?

Well-calibrated Dobson and Brewer direct-sun observations achieve repeatability near 1 percent and accuracy of roughly 1–2 percent. Ozonesondes provide profiles accurate to about 5–10 percent, with integrated columns agreeing with spectrophotometers to within 2–3 percent.

Which standards and guidelines govern these measurements?

Column measurements follow WMO/GAW operational handbooks for Dobson and Brewer instruments and NDACC protocols; surface ozone determination follows ISO 13964 and EN 14625, together with applicable national ambient air standards. Accredited third-party laboratories and national meteorological services apply these frameworks with documented traceability to primary references.

Why Choose ZKGX?

  • State-of-the-art analytical equipment
  • Highly qualified scientific team
  • Fast turnaround time
  • Competitive pricing